Effect of Dry Eye Treatment on Corneal Astigmatism Measurements
This prospective single-center study evaluates the effect of dry eye disease and short-term ocular surface treatment on the agreement and repeatability of corneal astigmatism measurements in patients with age-related cataract. Participants undergoing preoperative cataract evaluation are classified into a control group or a dry eye disease group according to ocular surface symptoms and objective tear film findings.
All participants undergo ocular surface assessment and repeated corneal measurements using three commonly used devices: IOLMaster 700, Pentacam AXL, and iTrace Prime. Measurements include keratometric corneal astigmatism, total corneal astigmatism, and vector components of astigmatism. Patients with dry eye disease receive 0.3% sodium hyaluronate eye drops four times daily for 2 weeks, after which ocular surface assessment and device measurements are repeated.
The primary objective is to determine whether dry eye disease reduces interdevice agreement and test-retest repeatability of corneal astigmatism measurements, and whether short-term ocular surface treatment improves measurement stability. Interdevice agreement is assessed using Bland-Altman analysis, and repeatability is evaluated using within-subject standard deviation, test-retest repeatability, and intraclass correlation coefficient.